Please use this identifier to cite or link to this item: http://dx.doi.org/10.18419/opus-4567
Authors: Faridian, Ahmad
Pedrini, Giancarlo
Osten, Wolfgang
Title: Opposed-view dark-field digital holographic microscopy
Issue Date: 2014
metadata.ubs.publikation.typ: Zeitschriftenartikel
metadata.ubs.publikation.source: Biomedical optics express 5 (2014), S. 728-736. URL http://dx.doi.org/10.1364/BOE.5.000728
URI: http://nbn-resolving.de/urn:nbn:de:bsz:93-opus-90892
http://elib.uni-stuttgart.de/handle/11682/4584
http://dx.doi.org/10.18419/opus-4567
Abstract: Scattering and absorption belong to the major problems in imaging the internal layers of a biological specimen. Due to the structural inhomogeneity of the specimen, the distribution of the structures in the upper layers of a given internal structure of interest is different from the lower layers that may result in different interception of scattered light, falling into the angular aperture of the microscope objective, from the object in each imaging view. Therefore, different spatial frequencies of the scattered light can be acquired from different (top and bottom) views. We have arranged an opposed-view dark-field digital holographic microscope (DHM) to collect the scattered light concurrently from both views with the aim to increase the contrast of internal structures and improve the signal-to-noise ratio. Implementing a DHM system gives the possibility to implement digital refocusing process and obtain multilayer images from each side without a depth scan of the object. The method is explained and the results are presented exemplary for a Drosophila embryo.
Appears in Collections:07 Fakultät Konstruktions-, Produktions- und Fahrzeugtechnik

Files in This Item:
File Description SizeFormat 
boe_5_3_728.pdf2,38 MBAdobe PDFView/Open


Items in OPUS are protected by copyright, with all rights reserved, unless otherwise indicated.