Titel: Improved reliability of impedance cardiography by new signal processing techniques
Erscheinungsdatum: 1988
Dokumentart: Konferenzbeitrag
Erschienen in: Proceedings of the 10th annual international conference of the IEEE Engineering in Medicine and Biology Society : frontiers of engineering and computing in health care. 3 : Computers and signal processing in medicine. Piscateway, 1988, S. 43-44
URI: http://nbn-resolving.de/urn:nbn:de:bsz:93-opus-54224
http://elib.uni-stuttgart.de/handle/11682/7260
http://dx.doi.org/10.18419/opus-7243
Zusammenfassung: This study used the impedance cardiogram (ICG), the phonocardiogram (pCG) and the electrocardiogram (ECG) to determine stroke volume and systolic time intervals. Comparisons between a fully automated PC-AT signal processing system and a computer-assisted visual detection processing system revealed comparable means, standard errors, correlations (Spearman r) and reliability (Chronbach's ex) coefficients on all relevant parameters. Other analyses confIrmed the utility of using Q onset to mark the beginning of ventricular systole, the B point of the ICG to denote the aortic opening and end diastole as the reference for determining the amplitude of dZ/dt max.